(818) 882-5000 [email protected] Northridge, CA 91324
AS9100D ISO 9001:2015

MGO-PAC Thermocouple Probes —Mineral Insulated Immersion Series

The Thermometrics MGO-PAC series comprises direct-immersion thermocouple probes with MgO-insulated, metal-sheathed construction for process temperature measurement in harsh industrial environments. Rigid, hermetically sealed, and available in all standard types and configurations.

MgO Insulated316SS / Inconel SheathType K/J/T/E/NGrounded / UngroundedLead Wire or Connector1-3 Day Lead Time
Model Series

MGO-PAC Probe Model Selection Guide

Select the configuration that matches your installation requirements. All models are available in Type K, J, T, E, or N. Custom immersion lengths, sheath ODs, and process connections manufactured to print.

ModelNameTerminationProcess ConnectionSheath ODTypical Application
TC-11MGO-PAC Immersion Probe — Lead Wire Flying leads, PTFE or fibreglass insulationNone (bare probe)1/16" to 1/4"OEM integration, portable measurement, lab use
TC-12MGO-PAC Probe — Miniature Connector Miniature TC connector (ANSI color-coded)None (bare probe)1/16" to 1/4"Instruments, analyzers, test & measurement equipment
TC-13Tube & Wire — Cut-to-Length Bare conductors, customer-specified lengthNone1/16" to 3/16"High-volume OEM, custom harness assemblies
TC-14MGO-PAC Probe — Threaded Fitting & Lead PTFE or fibreglass lead wire1/2" NPT threaded (standard); BSP available3/16" to 1/2"Pipe and vessel immersion, general process
TC-15Replacement MGO-PAC Element — Spring-Loaded Bare leads or miniature connectorHex head for industrial thermowell / assembly3/16" standardDrop-in replacement element for TC-70, TC-80 assemblies
Specifications

Technical Specifications

ParameterSpecification
TC TypesK, J, T, E, N (base metal); R, S, B (noble metal on request)
Junction StyleGrounded · Ungrounded (isolated) · Exposed
Sheath OD1/16" (1.6mm) · 1/8" (3.2mm) · 3/16" (4.8mm) · 1/4" (6.4mm) · custom
Sheath Materials316 SS (standard) · 304 SS · Inconel 600 · Inconel 625
InsulationMgO — high-purity grade, per ASTM E585
ToleranceASTM E230 Standard · Special Limits available
Insulation Resistance>100 MΩ at 20°C; >1 MΩ at max operating temp
Temp Range−200°C to 1260°C (type and sheath dependent)
Lead WirePTFE 24 AWG (standard) · Fibreglass · Ceramic-fiber for high-temp
Lead Length12", 24", 36" standard; custom lengths available
CalibrationNIST-traceable certificate available on request
CertificationsC of C standard; FAIR per AS9102 available

Junction Style Selection

  • GroundedFastest response; not isolated
  • UngroundedElectrically isolated from sheath
  • ExposedUltra-fast; gas/flow measurement only

Ordering Note

Specify: Model number · Thermocouple type · Sheath OD · Sheath material · Insertion length · Junction style · Lead wire type and length · Process connection (TC-14) · Quantity. Standard configurations ship in 1 business day. Call (818) 882-5000 for same-day availability.

Standard Lead Times

  • TC-11, TC-12, TC-13Same Day – 1 Day
  • TC-14 (with fitting)1 – 2 Days
  • TC-15 (replacement)1 Day
  • Custom OD / Special alloyQuote Basis
Key Features
01

MgO Mineral Insulation

High-purity magnesium oxide insulation maintains dielectric strength at operating temperature, compacts tightly during swaging, and allows the sheath to be bent without cracking the insulation layer.

02

Swaged Sheath Construction

MIMS cable is swaged to final OD during manufacture, cold-working the sheath and compressing the MgO fill to eliminate voids. The result is a rigid, vibration-resistant probe with no loose internal components.

03

Hermetic Tip Seal

Swaged closed-end tip or TIG-welded end cap provides a hermetic moisture barrier. Critical for applications involving process fluid pressure, wash-down environments, or steam sterilization.

04

All Standard TC Types

Type K for general purpose up to 1260°C. Type J for reducing atmospheres. Type T for cryogenic and food service. Type N for high-temperature stability. Noble metal Types R and S available for furnace calibration work.

05

ASTM E230 Lot-Certified Wire

All thermocouple wire is purchased to ASTM E230 tolerances with lot-test certifications. Special limits of error (SLE) — approximately half of standard tolerances — available on Type K, J, T, E, and N for tighter process control.

06

AS9100D Documentation

Certificate of Conformance standard on all orders. Material certifications, calibration certificates, and First Article Inspection Reports (AS9102) available for aerospace and defense programs. No minimum order quantity.

Related Products

Complete the Assembly

Ready to Specify Your Sensor?

Talk directly to the engineering team that builds your assemblies. No bots, no call centers.